Transcutaneous paCO2 Monitoring: Difference between revisions
From Guide to YKHC Medical Practices
mNo edit summary |
mNo edit summary |
||
Line 1: | Line 1: | ||
Transcutaneous CO<sub>2</sub> (TC-CO<sub>2</sub>) monitors function by measuring the CO<sub>2</sub> which diffuses out the skin. Modern TC-CO<sub>2</sub> monitors are a highly accurate, non-invasive method for estimating paCO<sub>2</sub>.<ref>Nassar BS, Schmidt GA. '''Estimating Arterial Partial Pressure of Carbon Dioxide in Ventilated Patients: How Valid Are Surrogate Measures?''' Ann Am Thorac Soc. 2017 Jun;14(6):1005-1014. doi: | Transcutaneous CO<sub>2</sub> (TC-CO<sub>2</sub>) monitors function by measuring the CO<sub>2</sub> which diffuses out the skin. Modern TC-CO<sub>2</sub> monitors are a highly accurate, non-invasive method for estimating paCO<sub>2</sub>.<ref>Nassar BS, Schmidt GA. [https://doi.org/10.1513/AnnalsATS.201701-034FR '''Estimating Arterial Partial Pressure of Carbon Dioxide in Ventilated Patients: How Valid Are Surrogate Measures?'''] Ann Am Thorac Soc. 2017 Jun;14(6):1005-1014. doi: 10.1513/AnnalsATS.201701-034FR. PMID: 28570147.</ref><br> | ||
<br> | <br> | ||
Revision as of 13:22, 13 March 2023
Transcutaneous CO2 (TC-CO2) monitors function by measuring the CO2 which diffuses out the skin. Modern TC-CO2 monitors are a highly accurate, non-invasive method for estimating paCO2.[1]
Primary Purpose
YKDRH obtained TC-CO2 monitors for the purpose of following paCO2 trends in patients receiving high-flow nasal cannula or non-invasive ventilation (i.e. CPAP/BiPAP). This was necessary because end-tidal CO2 monitoring is not compatible with these interventions.
Important Points
- Applicable for all age groups (including extremely premature neonates).
- Initial reading requires approximately five minutes to establish.
- TC-CO2 measures paCO2 trends; it DOES NOT measure instantaneous levels. Therefore, TC-CO2 is NOT an acceptable substitute for ET-CO2 in these situations:
- Confirmation of ET-tube placement.
- Respiratory monitoring during sedation.
- Any situation where apnea is a concern.
Device
YKDRH has acquired the Sentec Digital Monitoring System.
The Sentec devices provides monitoring of:
- Instantaneous oximetry
- Instantaneous heart rate
- Trending of paCO2
Resources
Instructional Videos:
- Neonatal Sensor Application for the Sentec Digital Transcutaneous Monitoring System
- Sensor Application Adult Cheek for the Sentec Digital Transcutaneous Monitoring System
REFERENCES
- ↑ Nassar BS, Schmidt GA. Estimating Arterial Partial Pressure of Carbon Dioxide in Ventilated Patients: How Valid Are Surrogate Measures? Ann Am Thorac Soc. 2017 Jun;14(6):1005-1014. doi: 10.1513/AnnalsATS.201701-034FR. PMID: 28570147.